Cochlear Implant Settings Update

Important update of settings for Cochlear implants, related to the detection of erosion and the possibility of dissolution of active implant electrodes.
Important update of settings for cochlear Nucleus 22, N24, CI500 and series cochlear implants Profile.

In letter from the firm Cochlear dated from 22 December 2017, sent to all US hospitals, reported that they found erosion, dissolution of active electrodes of implanted implants due to the use of bipolar stimulation.

In the above-mentioned letter, Cochlear recommends completely abandoning bipolar stimulation and reprogramming to all patients in a monopolar stimulation regimen.